Flowcheck™
Product No. 800501
Patent Applied For

Easy To Use

  • Fits any ½" HVE valve
  • Digital readout gage

Evaluates Open Suction

  • Use in flowing HVE valve
  • Flow tests vacuum station
  • Detects clogged vacuum lines

Accurate

  • Responds to small changes in flow or vacuum strength
  • High repeatability

Construction

  • Digital gage with ±0.25% full-scale accuracy
  • Precision machined, anodized aluminum

FLOWCHECK™ Uses

Check Treatment Room for minimal useful vacuum performance

  • Ensure vacuum system can support safe, efficient operation of high volume evacuators and nitrous oxide scavengers.
  • The Flowcheck™ provides a quantitative evaluation of performance by measuring flow in SCFM at an open high volume evacuator.

Specify size for proposed systems

  • Acceptable performance is indicated by a minimum flow at each open HVE of x SCFM as measured by a RAMVAC Flowcheck™
  • Sample Specification: The vacuum must provide acceptable performance for x HVE equivalents (1 HVE equivalent to 5 saliva ejectors or 5 small surgical aspirators or 4 nitrous oxide scavengers) in simultaneous and continuous operation.

Acceptance check - New equipment

  • Acceptable performance is indicated by a minimum flow at each open HVE of x SCFM as measured by a RAMVAC Flowcheck™
  • Sample Specification: Before acceptance, the vacuum must provide acceptable performance for x HVE equivalents (1 HVE equivalent to 5 saliva ejectors or 5 small surgical aspirators or 4 nitrous oxide scavengers) in simultaneous and continuous operation.

Performance check - Old equipment

  • Repeated testing can identify clogged lines and poorly cleaned solids separators and vacuum hoses.